Models and signal processing

openfrenz can turn the band's streams into numbers you care about: filter them, measure rhythms, find heartbeats, remove artifacts, and run a trained model on the result, on a recording or on a band as it streams. There are three ways in, all running the same Rust code:

You want to Use
Compute features in Python, for analysis or to train a model openfrenz.dsp
Run a model that is described entirely by its files openfrenz.Pipeline (Python) or frenz_model::Pipeline (Rust)
Write a model in Rust with any logic you like the Model trait in frenz-model

Because training and deployment share one implementation, a feature computed in Python for training is the same number, to rounding, as the one a deployed pipeline computes live.

Units

Stream values are raw device counts. The microvolt scale of the EEG is not known, so absolute powers are in squared counts; relative powers, coherence and coupling do not depend on the scale.

Blocks in Python

import numpy as np
import openfrenz
from openfrenz import dsp

session = openfrenz.export("recordings/FRENZA12-1795203836.sqlite")
t, eeg, segment = session.signal("eeg")          # eeg is (4, n) raw counts at 125 Hz

A signal is a NumPy array: (n,) for one channel or (channels, n). Channel order is that of the stream (openfrenz.SIGNALS).

Filters

clean = dsp.Filter(
    [{"type": "dc_block"},
     {"type": "notch", "hz": 60},
     {"type": "bandpass", "low": 0.5, "high": 45}],
    rate_hz=125, start="settled")

y = clean(eeg)                # filters these samples and remembers where it stopped
y2 = clean(more_eeg)          # continues exactly as if both had been filtered together
clean.reset()                 # start over, e.g. after a hole in the data
smooth = clean.filtfilt(eeg)  # forwards and backwards: no delay, needs the whole signal
Filter Settings Notes
lowpass high, order (2) Butterworth; -3 dB at high
highpass low, order (2) Butterworth; -3 dB at low
bandpass low, high, order (2) Butterworth; has 2 * order poles, as in SciPy
notch hz, q (30) q is the centre frequency over the notch's width
dc_block pole (0.99) removes a constant offset; a pole nearer 1 settles more slowly

Filters keep their state, which is what a live model needs: filtering a signal 2 seconds at a time gives exactly the numbers that filtering it whole gives. start="settled" begins as if the first sample had always been there, which avoids a large transient when the signal has an offset (raw EEG counts do); start="zero" (the default) begins from silence, as SciPy does. The numbers match SciPy's butter, iirnotch, sosfilt and sosfiltfilt.

Spectra and band power

freqs, density = dsp.welch(y, 125, segment=250)     # averaged over 2-second stretches
alpha = dsp.band_power(freqs, density, 8, 12)        # one value per channel
relative = alpha / dsp.band_power(freqs, density, 1, 45)
freqs, density = dsp.periodogram(y[:, :250], 125)    # a single stretch: noisier

Density is power per hertz, one-sided, with the mean removed and a Hann window, as SciPy's periodogram and welch. band_power sums the bins from the lower edge (included) to the upper (excluded), so adjacent bands do not share a bin.

Wavelets

amplitude, phase = dsp.morlet(y[0], 125, hz=10, cycles=6)

morlet follows a rhythm in time: the amplitude at hz for every sample, in signal units (a sine of amplitude 3 reads 3). More cycles separate neighbouring frequencies better and blur time more; 5 to 7 is usual. With 6 cycles a wavelet still sees about 1% of a rhythm an octave away. The first and last cycles / hz seconds read low, because the wavelet runs off the end of the signal.

Envelope and phase

beta = dsp.Filter({"type": "bandpass", "low": 20, "high": 30}, 125)
amplitude, phase = dsp.analytic(beta.filtfilt(y[0]))

analytic gives the amplitude envelope and the phase of a signal confined to a narrow band (as SciPy's hilbert), so filter first. Its first and last few cycles are distorted.

Coherence between channels

freqs, coh = dsp.coherence(y[2], y[0], 125, segment=125)   # forehead_left, forehead_right

Coherence is 0 for unrelated signals and 1 for a fixed linear relation, per frequency. It is computed from several stretches of signal: a single stretch always reads 1, and a handful read too high. Use at least 8 to 10 stretches (here, 5 seconds or more for 1-second stretches).

Coupling between rhythms

theta = dsp.Filter({"type": "bandpass", "low": 4, "high": 8}, 125)
gamma = dsp.Filter({"type": "bandpass", "low": 30, "high": 45}, 125)
_, phase = dsp.analytic(theta.filtfilt(x))
amplitude, _ = dsp.analytic(gamma.filtfilt(x))
c = dsp.coupling(phase[125:-125], amplitude[125:-125])      # leave out the distorted ends
c.modulation_index, c.mean_vector_length, c.preferred_phase

Phase-amplitude coupling asks whether a fast rhythm is larger at a particular phase of a slow one. Two measures are returned: Tort's modulation index (0 for none; a few hundredths is already strong) and a mean vector length scaled to lie between 0 and 1 whatever the signal's size. Both need many cycles of the slow rhythm, so several seconds of signal, and both respond to sharp artifacts such as blinks and muscle bursts, which have slow and fast parts locked together. Clean the signal first and treat a high value on a noisy stretch with suspicion.

Heartbeats from the pulse sensor

t, ppg, _ = session.signal("ppg")                 # (2, n) at 25 Hz
h = dsp.heart(ppg[0], 25)
h.rate_bpm, h.sdnn_ms, h.rmssd_ms, h.quality, h.beats

heart band-passes the pulse signal to 0.7–3.5 Hz, finds each beat as a rise, and times it between samples. It expects pulses that point up; pass -ppg if yours point down. The two pulse channels can differ a lot in how clean they are, so try each and keep the one with the higher quality.

Value Meaning
beats time of each beat, seconds from the start of the signal
rate_bpm beats per minute, from the mean interval
sdnn_ms standard deviation of the intervals between beats
rmssd_ms root mean square of the change from one interval to the next
quality share of intervals that look like heartbeats (plausible length, close to the median, between beats of ordinary size)

Distrust everything else when quality is below about 0.8: movement easily swamps the pulse. Variability needs a long stretch (a minute or more for SDNN), and the band's 25 Hz pulse signal times each beat to a few milliseconds at best, which adds that much to both variability measures. They are fine for comparing conditions and not a substitute for an ECG.

Removing artifacts

Artifact subspace reconstruction removes large, brief disturbances (blinks, jaw clenches, cable tugs) from multi-channel data. It learns from a quiet stretch how large each direction of the signal normally is; later, directions that are far larger than that are removed and rebuilt from the others.

clean = dsp.Filter([{"type": "dc_block"}, {"type": "bandpass", "low": 0.5, "high": 45}], 125, start="settled")
quiet = clean(eeg[:, :125 * 60])                       # a minute of quiet data, filtered
mixing, threshold = dsp.asr_calibrate(quiet, 125, cutoff=20)

asr = dsp.Asr(mixing, threshold)
cleaned = asr(clean(eeg[:, 125 * 60:]), 125)           # keeps its state: call it step after step
  • Calibrate on data filtered the same way as the data you will clean, and without drift or offset.
  • cutoff is how many times its usual spread a direction may exceed its usual size before it is removed. 20 to 30 is the recommended range (Chang et al. 2020); lower values remove more and start to remove brain signal.
  • A calibration belongs to its recording conditions. Different people, fits and days give different signals; a calibration made on one is not right for another.
  • Four channels limit what can be rebuilt. With few channels, a removed direction is rebuilt from little, so a disturbance is reduced, not erased: in our test a slow wave 25 times the signal's size on one channel is reduced about fivefold.

This is a compact form of the published method (Mullen et al. 2015; Kothe & Jung). It differs from the EEGLAB implementation in three ways: limits come from the median and spread of half-second sizes, not a fitted distribution; each half-second window is judged by its own data with no look-ahead, so there is no delay; and there is no built-in spectral weighting, so filter first.

Pipelines: models as data

A pipeline is a model written entirely in two files: config.json, which lists blocks and how they connect, and model.safetensors, which holds any stored numbers (normalisation, artifact calibration, network weights). It needs no code of its own, so any openfrenz installation can run it.

{
  "openfrenz_model": "pipeline",
  "name": "example", "version": "1",
  "step": { "stream": "eeg", "samples": 250 },
  "inputs": { "accel": { "samples": 100 } },
  "nodes": [
    { "name": "clean", "op": "filter", "in": "eeg", "start": "settled",
      "steps": [{ "type": "notch", "hz": 60 }, { "type": "bandpass", "low": 0.5, "high": 45 }] },
    { "name": "power", "op": "band_power", "in": "clean",
      "bands": [[4, 8], [8, 12], [12, 30]], "relative_to": [1, 45] },
    { "name": "coh", "op": "coherence", "in": "clean",
      "pairs": [["forehead_left", "forehead_right"]], "bands": [[8, 12]], "average_steps": 5 },
    { "name": "move", "op": "summary", "in": "accel", "stats": ["sd"] },
    { "name": "x", "op": "concat", "in": ["power", "coh", "move"] },
    { "name": "z", "op": "normalize", "in": "x", "weights": "norm" },
    { "name": "recent", "op": "history", "in": "z", "steps": 5 },
    { "name": "net", "op": "network", "in": "recent", "names": ["score"],
      "layers": [{ "type": "lstm", "weights": "rnn" }, { "type": "last" },
                 { "type": "dense", "weights": "head" }, { "type": "sigmoid" }] }
  ],
  "outputs": ["net"]
}

How to read it:

  • step names the stream that sets the pace and how many of its samples make one step (here 2 seconds of EEG). inputs lists the other streams and how many samples of each belong to a step; see Streams a model can do without for the other forms.
  • nodes run in order. Each has a name, the block it is (op), what it reads (in: a stream or an earlier node, or a list of them) and the block's settings.
  • outputs names the nodes whose values the model returns.
  • Stored numbers are named by a node's weights, a prefix into model.safetensors.

Three kinds of thing flow between nodes: a signal (channels by samples), features (named numbers describing one step) and a sequence (the features of several steps).

HyperStudy Bridge draws a pipeline as a graph and edits it there; what it saves is this same config.json.

Run a pipeline

import openfrenz

model = openfrenz.Pipeline("my-model")          # a folder, or "owner/name" on Hugging Face
model.inputs                                    # {'eeg': 250, 'accel': 100}
model.outputs                                   # ('score',)

scores = model.run("recordings/session.sqlite") # DataFrame: t and one column per output

with openfrenz.connect("FRENZA12") as band:
    for out in model.stream(band):              # {'t': ..., 'score': ...} every step
        print(out)

out = model.step({"eeg": eeg_block, "accel": accel_block})   # one step by hand; None while warming up
use frenz_model::{Live, Pipeline, run_recording};

let model = Pipeline::from_dir(Path::new("my-model"))?;
let run = run_recording(model, Path::new("session.sqlite"))?;       // run.outputs: Vec<Output { t, values }>

let mut live = Live::new(Pipeline::from_dir(Path::new("my-model"))?, &band, true)?;
while let Some(output) = live.next().await? { /* ... */ }

run scores a recording as fast as the model computes. stream gives one result per step as the band delivers data. Both restart the model wherever data is missing (a new segment, a hole inside a step), so nothing is carried across a gap, and both leave out repeated and frozen packets. Live, gaps are recognised by when packets arrive: replay a recording at its recorded pace to see what a live run would have given, and use run to score it quickly.

A step is skipped when a stream the model requires is missing from it. Skipped steps are counted by the stream that was missing, so a model that stays silent can say why:

stream = model.stream(band)
out = stream.read(timeout=10)        # None if no step completed in 10 seconds
stream.missed                        # {'ppg': 4}: four steps skipped for want of the pulse stream
scores.attrs["missed"]               # the same count for a scored recording

In Rust the counts are Live::missed() and the missed field of what run_recording returns. The pulse sensor, for example, sends nothing until the band is on the head.

Streams a model can do without

By default every stream in inputs is required, and a step without one is skipped and restarts the model. Two other forms avoid that:

"inputs": {
  "accel":   { "samples": 100 },
  "ppg":     { "samples": 50, "required": false },
  "battery": { "arrived": true }
}
  • "required": false uses the stream when it is there. While it is missing, every feature made from it is not-a-number and the rest of the model carries on, without a restart. Blocks that gather several steps of it start gathering again when it returns, and give not-a-number until they have enough. Use this for streams that drop out on their own, such as the pulse sensor, which freezes for a second or two now and then.
  • "arrived": true is for streams without a steady rate (heart_rate, spo2, battery, contact): a step gets whatever readings arrived during it, often none. summary of nothing is not-a-number.

Not-a-number travels through concat, normalize, history and network, so keep optional features out of a network's inputs unless the network was trained to be given a stand-in value.

A pipeline is checked when it is loaded, by running it on silence. A block given the wrong kind or amount of input is reported then, with the node's name:

config.json: node z: was stored for 35 features and given 18

Blocks

Block Reads → gives Settings (defaults)
select signal → signal channels: names to keep, in order
mix signal → signal channels: list of {"name": ..., "weights": {channel: weight}}. Each new channel is a weighted sum, for example a difference between two electrodes
filter signal → signal steps: filters as in Filters; start: zero, settled, or a number (that value, held for ever) (zero); zero_phase: filter each step both ways, without state (false)
asr signal → signal weights: prefix of mixing and threshold; max_removed (0.66); window_s (0.5); weighting: {"b": [...], "a": [...]}; average_windows (1); blend_offset: 1 or 0 (1); before: filters run inside the node; before_start; passes (1). See Matching another cleaner
band_power signal → features bands: list of [low, high]; relative_to: [low, high] to divide by; log (false); segment: average over stretches of that many samples; measure: power or magnitude (power); integrate: sum or simpson (sum); window: hann or none (hann); remove_mean (true). See Shares, ratios and averages
wavelet signal → features frequencies: list of Hz; cycles (6); trim_s: seconds left out at each end (0); log (false). Gives the mean amplitude over the step
envelope signal → features band: [low, high]; order (2); stats (["mean", "sd"])
summary signal → features stats: any of mean, sd, log_sd, range, largest_step (["mean", "sd"])
quality signal → features limit: the largest value the sensor reports. Gives the shares of repeated, clipped and invalid samples
coherence signal → features pairs: list of [channel, channel]; bands; segment (125); average_steps (1)
phase_amplitude signal → features phase_band, amplitude_band; order (2); trim_s (0); stats: any of amplitude_mean, amplitude_sd, phase_mean, phase_sd, correlation (all)
coupling signal → features phase_band, amplitude_band; order (2); bins (18); trim_s (0); average_steps (1)
heart signal → features channel (the first); invert (false); window_steps (1). Gives rate, SDNN, RMSSD and quality
peaks signal → features threshold (4); back_s (2); ahead_s (0.25); min_gap_s (0.25); scale_half_life_s (120); warmup_s (5). Gives peaks, peak_time and peak_size; see Moments
jumps signal → features threshold (6); leak_s (5); min_gap_s (0.12); near_s (0.048); far_s (0.152); scale_half_life_s (120); warmup_s (5). Gives position and jumps
tilt signal → features counts_per_g (16384). Reads the accelerometer's x, y, z; gives pitch_deg, roll_deg, from_upright_deg and g; see Head posture and movement
ratios features → features pairs: list of [measure, measure]; log (false)
average features → features channels: which to average (all); prefix of the averaged features (mean)
concat features, ... → features
normalize features → features weights: prefix of stored center and scale; or half_life_steps and warmup_steps (2) to track them as steps pass
history features → sequence steps: how many to keep
network features or sequence → features layers; names: one per output value

Things to know:

  • State. filter, asr, coherence, coupling, heart, peaks, jumps, history and a running normalize carry state from step to step, and all of it is cleared at a reset.
  • Warm-up. A block that needs several steps (average_steps, window_steps, steps, warmup_steps) gives nothing until it has them, and neither does anything that reads it. The model's first output comes when every block is ready.
  • Averaging over steps. Coherence and coupling cannot be measured in one short step. With average_steps: 5 and 2-second steps they describe the last 10 seconds, updated every step.
  • Feature names are built from the channel and the measure, for example forehead_left:power:8-12, forehead_left~forehead_right:coherence:8-12, x:sd. A model's outputs lists them.
  • Network layers are dense and lstm (each with weights, and dense optionally with an activation), last, mean and flatten to turn a sequence into one row, and relu, elu, tanh, sigmoid and gelu. Arrays are laid out as PyTorch stores them: prefix.weight and prefix.bias for dense; prefix.weight_ih_l0, prefix.weight_hh_l0, prefix.bias_ih_l0 and prefix.bias_hh_l0 for lstm. The network must end in a single row.

Moments: peaks and jumps

Two blocks find things that happen at a moment instead of describing a whole step. Both work sample by sample, so the result does not depend on how long the steps are.

peaks finds peaks that stand out from their surroundings: local maxima whose height above the higher of the two lowest points nearby (within back_s before and ahead_s after) is at least threshold times the signal's ordinary size. The ordinary size is a running median of the signal's absolute value, so the peaks themselves do not raise it; it follows slow changes over about scale_half_life_s. Of peaks closer together than min_gap_s, only the highest counts. Each step gives, per channel:

Feature Meaning
peaks how many peaks became certain during this step
peak_time when the largest of them was, in seconds counted back from the step's last sample (zero or negative; not-a-number if none)
peak_size how far it stood out, in units of the ordinary size (not-a-number if none)

A peak is reported ahead_s to ahead_s + min_gap_s after it happened, so it often belongs to an earlier step than the one that reports it. Its time is the step's t plus peak_time:

out = model.run("recordings/session.sqlite")
found = out[out["blink:peaks"] > 0]
blink_times = found["t"] + found["blink:peak_time"]

jumps follows a level that moves in abrupt steps while the sensor's baseline wanders. A change between two samples counts as a jump when it is threshold times the ordinary change; its height is the mean level from near_s to far_s after it minus the same before it. Heights are added to a position that fades back to zero, losing half of itself every leak_s. Slow wander never reaches the position, and a level that changes and stays changed is still seen, which a high-pass filter would lose. jumps counts the jumps added during the step, each far_s after it happened.

Neither block detects anything for the first warmup_s, while it learns the ordinary size.

Matching another cleaner

The asr block's options let a pipeline reproduce artifact subspace reconstruction as other software runs it, sample for sample:

  • weighting filters the data before its size is judged (the cleaned data itself is not filtered), as EEGLAB does with its spectral weighting. Give the filter's coefficients, b over a; it starts settled on the first sample.
  • average_windows judges each window by the average covariance of it and the windows before it.
  • blend_offset sets where the ease from one window's reconstruction into the next starts: 1 (EEGLAB) or 0.
  • before runs a filter chain inside the node, before the reconstruction, starting as before_start says. On its own it is the same as a filter node in front.
  • passes runs that chain and the reconstruction several times over each step and keeps the first result, so both see each step that many times. Some pipelines were trained this way, and their models need it to give the same numbers.
  • max_removed of 0.75 with four channels always keeps one direction.

Shares, ratios and averages

Three blocks turn per-channel spectra into the kind of numbers a model is usually given:

  • band_power gives power per band by default. With "measure": "magnitude" it gives the size of the Fourier transform instead of its square, over the whole step without averaging. "integrate": "simpson" integrates each band with Simpson's rule over the bins nearest its edges, both included, instead of summing bins. With relative_to, each band is a share of that range.
  • ratios divides one measure by another for every channel that has both. The pair ["magnitude:4-8", "magnitude:8-12"] turns forehead_left:magnitude:4-8 and forehead_left:magnitude:8-12 into forehead_left:magnitude:4-8/magnitude:8-12, and the same for each other channel. Any two measures can be paired, from any blocks, as long as both reach the ratios node.
  • average averages each measure over channels: forehead_left:x and forehead_right:x become mean:x (prefix sets another word than mean). Averaging after ratios gives the mean of the ratios; averaging the signals first would not.
{ "name": "band", "op": "filter", "in": "front", "zero_phase": true,
  "steps": [{ "type": "bandpass", "low": 2, "high": 45, "order": 1 }] },
{ "name": "shares", "op": "band_power", "in": "band", "bands": [[4, 8], [8, 12]], "relative_to": [0, 62],
  "measure": "magnitude", "integrate": "simpson", "window": "none" },
{ "name": "ratio", "op": "ratios", "in": "shares", "pairs": [["magnitude:4-8", "magnitude:8-12"]] },
{ "name": "mean", "op": "average", "in": ["ratio"] }

phase_amplitude relates the phase of one band to the amplitude envelope of another within each step: the envelope's mean and spread, the phase's mean and spread after removing its jumps of 2π, and the correlation between the two. Both bands are filtered both ways, so the step's ends are slightly distorted; trim_s leaves them out. coupling measures something different: how the envelope's size depends on the phase, over several steps.

Write a pipeline from Python

Train with features from openfrenz.dsp, then save the same steps as a pipeline. With PyTorch and the safetensors package:

import json
import torch
from safetensors.torch import save_file

# rnn = torch.nn.LSTM(features, 16, batch_first=True); head = torch.nn.Linear(16, 1)  ... trained
weights = {f"rnn.{k}": v for k, v in rnn.state_dict().items()}
weights |= {f"head.{k}": v for k, v in head.state_dict().items()}
weights |= {"norm.center": torch.tensor(train_mean), "norm.scale": torch.tensor(train_sd)}
save_file(weights, "my-model/model.safetensors")

with open("my-model/config.json", "w") as f:
    json.dump(config, f, indent=1)                  # the dict shown above

openfrenz.Pipeline("my-model")                      # loads and checks it

An artifact calibration is stored the same way, as prefix.mixing and prefix.threshold from dsp.asr_calibrate. Compare the pipeline's outputs with your training code on a few steps before relying on it: model.step(...) makes that a few lines.

Share a model on Hugging Face

Upload the folder (config.json, model.safetensors, a README) to a model repository, and load it by name:

model = openfrenz.Pipeline("owner/name")            # or "owner/name@v1.0", "owner/name@<commit>"

Files are downloaded once per commit into ~/.cache/openfrenz (set OPENFRENZ_CACHE to move it); a branch name is looked up each time to see which commit it points to, and when Hugging Face cannot be reached the commit it pointed to last time is used, so a model fetched once also works offline.

Private repositories need a token from an account with access. Give it as token, or let openfrenz find one in HF_TOKEN or the Hugging Face command-line login:

model = openfrenz.Pipeline("lab/private-model", token=my_token)

A repository that cannot be read (private without a token with access, or not there) raises an error that says which; HF_ENDPOINT points openfrenz at a mirror. The files hold numbers and settings only and are never executed, and the sizes a config.json can ask for are bounded (samples per step, filter order, steps kept, and so on), so a model from an unknown source can be slow or wrong but cannot run code. Model files over 512 MB are not downloaded.

Models that come with openfrenz

Five pipelines are built in and are loaded by name. None needs stored numbers. openfrenz.MODELS lists them.

Name Reads Gives, per step
eog eeg eye signals: blinks and horizontal gaze, every 0.2 s
brow-emg eeg muscle activity of the forehead, every 0.2 s
heart ppg heart rate, its variability and a quality score over the last 30 s, every 5 s
head accel, gyro how the head is held and how much it moves and turns, every second
eeg-power eeg power in the delta, theta, alpha, beta and gamma bands of each electrode and their average, every second

Live, as the band streams:

model = openfrenz.Pipeline("eog")
with openfrenz.connect("FRENZA12") as band:
    for out in model.stream(band):
        if out["blink:peaks"]:
            print("blink at", out["t"] + out["blink:peak_time"])

In an exported session, as one more table beside the streams:

session = openfrenz.export("recordings/session.sqlite", models=["eog", "brow-emg", "heart"])
session["eog"]                              # DataFrame: t and one column per output
frenz export recordings/session.sqlite --model eog --model brow-emg --model heart

models and --model also take a folder or a Hugging Face repository, so your own pipelines are exported the same way. The recording itself is never changed: it stays the raw record, and a session can be exported again whenever a model improves.

Which model made a table

These models will change as more data is collected. Each model table in session.json records where its numbers came from:

session.meta["tables"]["eog"]["model"]
# {'name': 'eog', 'version': '0.1', 'source': 'eog',
#  'fingerprint': '<16 hex digits>', 'runner': 'openfrenz frenz-model 0.3.0',
#  'outputs': [...], 'missed': {'eeg': 1}}
  • version is what the model's config.json calls itself.
  • fingerprint is computed from the model's files and changes whenever they do, even if nobody remembered to change the version.
  • runner is the openfrenz version that ran it; session.meta["exporter"] is the one that exported the streams.

To find sessions made with an older model, compare with the model in hand:

current = openfrenz.Pipeline("eog").fingerprint
stale = [p for p in openfrenz.find_sessions("sessions")
         if openfrenz.load_session(p).meta["tables"].get("eog", {}).get("model", {}).get("fingerprint") != current]

Eye and brow signals

The eyes and the forehead muscles are read from the same electrodes in different frequency ranges. They are two models, so that each can change on its own. Both give a row every 0.2 s.

Model Output From Meaning
eog blink:peaks, blink:peak_time, blink:peak_size forehead pads minus side combs, 0.5-8 Hz, peaks blinks and longer eye closures
eog gaze_horizontal:position, gaze_horizontal:jumps left minus right forehead pad, below 8 Hz, jumps where the eyes point, left to right, in device counts; fades to zero within seconds
brow-emg brow:log_sd left minus right forehead pad, 20-58 Hz, summary how hard the forehead muscles are working (natural logarithm)

These outputs are experimental. They were checked against face video analysed with py-feat, on two ten-minute sessions of one person on consecutive days:

Output Check Day 1 Day 2
Blinks F1 against blinks seen on video, within 0.2 s 0.49 0.70
Horizontal gaze correlation with gaze on video, during cued looks left and right 0.89 0.69
Brow correlation with brow raise on video, whole session 0.59 0.56

Blink times were within 15 ms of the video on both days. Version 0.2 of eog lowered the blink threshold from 4 to 3: on a third day, when the side combs were about 45 times noisier than the forehead pads, version 0.1 found 9 of 15 cued natural blinks and 0.2 finds all 15. The cost is more false detections on quieter days (F1 0.50 to 0.49 and 0.76 to 0.70 on the two video days). What to keep in mind:

  • Gaze is readable only while the face is still. Over a whole session, including posed expressions, its correlation with video is near zero: one large movement of the forehead adds a jump far bigger than any eye movement. It gives direction and relative size, not degrees.
  • Vertical gaze is not given. The band's electrodes do not separate it.
  • Brow level is relative. Compare it with a stretch of rest from the same session, or follow it with a running normalize.
  • Blink detection depends on the side combs. The blink signal is the forehead pads minus the combs, and how noisy the combs are varies with fit from day to day.
  • One person and two days cannot show how these hold across people, hair or fit.

EEG power bands

eeg-power filters each electrode (60 Hz notch, 1-45 Hz band-pass) and gives, every second, the natural logarithm of its power in 1-4 (delta), 4-8 (theta), 8-13 (alpha), 13-30 (beta) and 30-45 Hz (gamma), named like forehead_left:power:8-13, plus each band averaged over the four electrodes (mean:power:8-13).

On this band most of what is above about 20 Hz is muscle, not brain (see the eye and brow findings), so read beta and especially gamma as facial tension unless the face is still. Power is in squared device counts: compare it within a session, not between bands or people.

Heart rate and its variability

heart runs the heart block on one pulse channel (tag3) over the last 30 s and gives heart:rate_bpm, heart:sdnn_ms, heart:rmssd_ms and heart:quality every 5 s.

Also experimental, and not checked against an ECG. On the same two sessions quality was above 0.8 for 96% and 69% of steps, with a median rate of 72 and 71 beats a minute; the band's own heart-rate stream, which reported a value in only 2% and 10% of its readings, averaged 78 and 73. Keep only steps with quality above about 0.8. The other pulse channel (tag2) was unusable in both sessions. Thirty seconds is short for variability, and 25 Hz limits how finely each beat is timed; see the block's notes.

Head posture and movement

head reads the motion sensors and gives, every second:

Output Meaning
tilt:pitch_deg how far the head leans forward (positive, chin down) or back, from the direction of gravity
tilt:roll_deg how far it leans toward the wearer's left shoulder (positive) or right
tilt:from_upright_deg angle between the top of the head and straight up: near 0 upright, near 90 lying down
tilt:g size of the accelerometer reading in g: 1 when still
accel_down:sd, accel_left:sd, accel_forward:sd how much the head was jolted along each axis during the second (above 0.5 Hz, in counts)
turn_shake:sd, turn_nod:sd, turn_side:sd how much it turned about the vertical axis (a shake), the ear-to-ear axis (a nod) and the nose axis (ear toward shoulder), in gyroscope counts

What these can and cannot say:

  • Upright or lying down, and on which side. Lying on the back gives a pitch near -90, face down near +90, on the right side a roll near -90, on the left near +90.
  • Sitting and standing look the same: the head is upright in both. Telling them apart needs more than how the head is held, for example the moment of standing up.
  • The band sits at an angle on the forehead, and not the same angle every time: sitting upright read 12 and 23 degrees from upright on two days. Compare with a known upright stretch from the same session before reading small leans.
  • Nods and shakes are not counted yet. The turn outputs say how much the head turned about each axis, which is what published detectors of nods and shakes build on, but no recording with cued nods and shakes has been made to set and check a detector. The sensor's y and z axes are also rotated about 30 degrees from the band's, so nods and sideways tilts are not cleanly separated; shakes, about the vertical axis, are.

The angles follow from the sensor's axes and gravity and are tested on made-up readings. They have not been compared with a measured head angle.

Models in code

A pipeline covers models built from the standard blocks. For anything else, write the model in Rust against the Model trait: it declares the streams it reads and is called once per step. Such a model can read any stream the band sends, including the slow ones (heart_rate, spo2, battery, contact), can mark a stream as preferred, and can decide for itself what to do when one is absent. See Run a model in the Rust guide.

Limits

  • Pipelines run on .sqlite recordings and on bands, not yet on exported session folders.
  • The blocks are tested against SciPy, NumPy and PyTorch on synthetic signals. Artifact reconstruction and heartbeat detection have not been validated against reference recordings (simultaneous research-grade EEG or ECG).
  • Running a pipeline live has been checked on replayed recordings and, briefly, on one band.