FRENZ brainband: hardware and Bluetooth interface

What this project knows about the band, from recordings and bench tests with one unit (firmware X8.2.2_1.1.19, GATT firmware revision 20.3). Each statement is tagged:

  • [V] verified by measurement on hardware
  • [I] inferred from indirect evidence; treat as a working assumption

Corrections and results from other units or firmware versions are welcome. All multi-byte integers are big-endian.

Sensors and contacts

Part Details
EEG 4 channels, 125 Hz, signed 24-bit ADC counts [V]. Microvolt scale not yet known.
Forehead pads 4 dry pads. Two are recording sites: left of centre and right of centre (beside the optical sensor) [V]. The centre pad behaves like the reference and the remaining pad is presumably bias/ground [I].
Side combs 2 soft dry comb electrodes, one per side of the head, both recording sites [V].
Optical sensor PPG, 3 tagged channels at ~25 Hz, of which one is always zero [V]. Which LED each tag is: unknown.
Motion 3-axis accelerometer and 3-axis gyroscope, 50 Hz each, signed 16-bit counts [V].
Derived Heart rate and SpO2 computed on the band, ~1 Hz [V].
Status Battery percentage, electrode contact quality (0–3) [V].
Audio Bone-conduction speakers and an audio chipset are present; not characterized.

EEG channel map [V]

Packets carry six channel slots; four carry signal.

Slot Name in this project Electrode Typical content
0 forehead_right right forehead pad blinks, jaw EMG
1 comb_left left side comb ~10 Hz alpha with eyes closed; large DC offset; strong mains pickup
2 — unused, always 0
3 forehead_left left forehead pad blinks (strongest), jaw EMG
4 comb_right right side comb as comb_left
5 — unused, always 0

The map comes from pressing on the band over each contact in turn (the responding channel was 13–72× stronger than the next) and from eyes-open/eyes-closed recordings (alpha power ×1.6–1.8 on the combs).

Motion sensor frame and scale [V]

  • Axes: +x down (toward the chin when worn), +y toward the wearer's left, +z forward. The y/z axes are rotated about 30° around x relative to the band.
  • Accelerometer: ±2 g, 16-bit, ≈ 16 384 counts/g. A six-orientation static calibration of one unit gave per-axis scales within 1.3 % of nominal and biases of a few hundred counts; calibrate each unit.
  • Gyroscope: small at-rest bias (tens to ~100 counts). Scale is probably 65.5 counts per deg/s (±500 deg/s): hand-made quarter turns gave 50–62 [I]; a controlled rotation is needed to confirm.

Bluetooth LE interface

Band-specific services use the UUID suffix -0000-ffff-ff45-415241424c45. No pairing key or payload encryption is involved [V].

UUID prefix Properties Role
45420101 write without response commands
45420102 notify command responses
45420103 notify status messages
45420201 notify EEG
45420202 notify accelerometer
45420203 notify PPG
45420204 notify heart rate
45420205 notify SpO2
45420206 notify battery
45420207 notify gyroscope
45420208 notify contact quality

The band also exposes Google Fast Pair (fe2c), audio-chipset services and the standard battery (180f) and firmware-revision (2a26) characteristics. Subscribe only to the 4542… characteristics: subscribing to every notifying characteristic on the device dropped the connection [V].

Commands and responses

A response is [opcode, status, payload…] with status 00 for success [V].

Command Bytes Response payload
Firmware version 01 ASCII string
Battery 02 percent, u8
Stream control 03 <mask> 01 start, 03 ff 00 stop ack. Mask bits [V]: 01 EEG, 02 accelerometer + gyroscope, 10 contact, heart rate and SpO2. ff starts everything. The PPG bit is not yet identified (PPG takes ~6 s to start).
Serial number 06 ASCII
Time sync 07 + Unix seconds (u32) ack; packet timestamps become Unix seconds
Device name 09 ASCII
Read device time 0b u32 seconds
Device ID 0c ASCII (the Bluetooth address)
Session notice 0e 00 on connecting, 0e 02 after the stream start, 0e 03 at end ack. Without 0e 02 the band powers itself off 10 minutes after the stream start [V]; other effects not characterized
Periodic heart rate 10 01 ack; HR/SpO2 change from every 5 s to every 1 s
Prompt volume 36 + u8 ack
IMU calibration report f8 3 × u8 std, 3 × u16 diff; exact meaning [I]

Start sequence used by this project [V]: firmware version, time sync, prompt volume 0, periodic heart rate, session notice, stream start, IMU calibration report, session-running notice (0e 02). Stop: stream stop, session-end notice.

The device has further opcodes (audio, alarms, stimulation, file transfer, firmware update, test modes, reset, rename). They are intentionally not implemented here until characterized on a bench.

Stream payloads

Timestamps are whole seconds: seconds since power-on until a time sync, Unix seconds afterwards [V].

Stream Layout
EEG 178 bytes: sequence u8, type u8 (= 6), seconds u32, then 9 samples × (6 × i24 + flag u8), trailer u8. 13.9 packets/s (125.0 samples/s measured). Flag is 1 on ~4 % of samples and the trailer varies 0–16; meanings unknown.
Accelerometer, gyroscope 158 bytes: seconds u32, 25 × (x, y, z as i16), 4 trailing bytes (mostly zero; meaning unknown). 2 packets/s.
PPG seconds u32, then 12 or 13 samples × three 24-bit words, then 2 zero bytes (114 or 123 bytes). Each word: tag in bits 23–19 (1, 2, 3), unsigned 19-bit value in bits 18–0. 2 packets/s.
Heart rate, SpO2 6 bytes: seconds u32, value u8 (beats per minute or percent), then a byte that is 1 when the band has no reading (value is then 0) and 0 otherwise [V].
Battery 5 bytes: seconds u32, percent u8. Every 5 s.
Contact 5 bytes: seconds u32, state u8 (0–3). Every 2 s.
Status variable. 0e 00 00 00 00 about once a second. Every 60 s a block: 07 12 ff ee + u16 that tracks battery voltage in mV (4039 → 4010 as charge fell 90 → 87 %) [I], 07 12 aa aa + u16 (constant 37), 07 12 ff ff + u16 (44–62, unknown), 07 11. 07 db, 07 d0 are sent when the band is about to power off, whether by timeout or by its power button [V].

Behaviour worth knowing

  • Streaming state persists across connections [V]. After a session that sent the stop command, no EEG/motion/PPG/contact data arrives until stream start is sent again (heart rate, SpO2, battery and status still do). If the previous session did not stop the streams, data flows as soon as the characteristics are subscribed. Always send stream start.
  • The EEG sequence counter restarts at 0 when stream start arrives while streams are running [V].
  • Start-up artifacts [V]: about 0.8 s after every stream start, one EEG packet repeats the previous packet's samples exactly while the sequence counter still advances. Once (1 of 5 sessions) a block was repeated three times, the last copies arriving as frames without the leading sequence byte.
  • Contact quality settles slowly [V]: after a stream start it reads 0, 0, 1, 1, 3 at two-second intervals. 3 is good contact; it flickers to 2 with marginal contact and drops to 0 when the band is taken off.
  • PPG can freeze [V]: with marginal contact the PPG output held one constant value for 2–5 s and the band sent extra PPG packets per second during the freeze. PPG also switches off ~4 s after the band is removed.
  • Automatic power-off [V, one observation each]: while streaming, the band powered off ~33 s after being taken off (putting it back on within ~26 s kept it on); while not streaming it powered off ~11 minutes after the last disconnect.
  • Ten-minute power-off without the session-running notice [V]: when 0e 02 was not sent, the band sent the power-off warning 600.3 s after the stream start and dropped the link at 605 s, with good contact throughout (4 of 4 sessions). With 0e 02 sent it streamed for 13 minutes and stayed on (2 of 2 sessions). Longer sessions have not been tried.
  • Delivery jitter [V]: Bluetooth delivery delay beyond the minimum had a 95th percentile of 80–100 ms for EEG and motion and ~250–300 ms for PPG on macOS. Estimate sample times from sample counts, not arrival times (see frenz-export).
  • Streams are aligned [V]: taps on the band appear in the accelerometer and in the EEG within ±20 ms (one motion sample) of each other in exported time. EEG packets are stamped ~40 ms later than motion packets on the band's own whole-second clock, which is a labelling difference, not a delay.
  • Measured sample rates [V]: EEG 125.00 Hz, motion 50.00 Hz, PPG 25.00 Hz, each within 0.01 Hz over three minutes.

Open questions

  • EEG microvolt scale.
  • The four trailing bytes of motion packets; gyroscope scale.
  • Which optical channel is which LED, and why one is zero.
  • The PPG stream-control bit, the remaining status values and the IMU calibration report.
  • The EEG per-sample flag clusters in the last packet of each device second and the trailer is an 8-bit mask; both look timing-related, exact meaning unknown.
  • Behaviour on other firmware versions, sessions longer than an hour, and several bands at once.